DOTX-45 Dissolved Oxygen System Part 5 Sensor Assembly
DOTX-45 25 -
6. Slowly screw the chamber onto the sensor body. A small amount of
electrolyte will run out of the hole from which the vent screw was removed.
Place a paper towel around the sensor to absorb the electrolyte overflow.
The electrolyte is harmless and will not irritate skin. Tighten the chamber until
the o-ring at the top of the chamber is compressed. Once again, do not use
tools to tighten.
7. Shake excess electrolyte from the vent hole on the side of the sensor and
replace the vent screw.
The sensor is now ready for operation. The membrane should be stretched
tightly across the tip of the sensor.
CAUTION: When handling the assembled sensor, do not set the sensor on its
tip or damage to the membrane will result. Severe impacts on the tip
of the sensor from dropping or other misuse may cause permanent
damage to the sensor.